Sensata-Tech-DAT72405SU
| Part # | Sensata-Tech-DAT72405SU |
| Download | Sensata-Tech-DAT72405SU |
|
DA(UL) SERIES
UL APPROVED*, NORMALLY OPEN, HIGH VOLTAGE RELAYS – 10KV, 7.5KV & 5KV
The DA(UL) series is a UL* approved very high isolation voltages (up
to 10kV) are achieved through the use of high vacuum reed switches
with either rhodium or tungsten contacts and make these relays suitable
for high reliability applications, such as cardiac defibrillators, test
equipment and high voltage power supplies.
A choice of 5k
V
,
7
.5kV and 10kV isolation voltages is available.
The rhodium contact relays have low contact resistance, while the
tungsten contact relays can switch higher voltages.
PCB or panel mount, via nylon studs, versions are available.
Connection options, for the H
V
,
i
nclude PCB, solder turret (wire wrap),
flying lead and 0.25” spade terminals.
Features
• Choice of 10k
V
,
7
.5kV or 5kV Isolation
• Low contact resistance
• PCB or panel mount
• HV connections via flying leads, solder turret (wire wrap),
or 1/4” spade terminals
• Excellent AC characteristics
SPECIFICATIONS
Units
Condition
10kV
7.5kV
5kV
Contact
N/O (normally open)
Contact Form
Rhodium Tungsten Rhodium Tungsten Rhodium Tungsten
Contact Material
kV
W
V
DC or AC peak
10
50
1000
3
10
50
7000
2
7.5
50
7.5
50
5
50
1000
3
5
50
3500
2
Isolation Across Contacts
Switching Power Max
Switching Voltage Max.
Switching Current Max.
Carry Current Max
DC or AC peak
DC or AC peak
DC or AC peak
1000
3
5000
2
A
A
4
3
4
3
4
3
coil to screen
grounded
pF
<0.2
<0.2
<0.2
<0.2
<0.2
<0.2
Capacitance Across Contacts
Lifetime Operations
dry switching
10
9
10
6
10
9
10
6
10
9
10
6
10
9
10
6
10
9
10
6
10
9
10
6
50W switching
50 (15)
10
10
250 (100)
(10
13
)
50 (15)
10
10
250 (100)
(10
13
)
50 (15)
10
10
250 (100)
(10
13
)
m
Ω
max (typical)
Ω
min (typical)
Contact Resistance
Insulation Resistance
Page 1
Copyright © 2022 Sensata Technologies, Inc.
Units
V
Condition
DC
5V
3.7
0.5
3.0
2.0
28
12V
9
24V
20
Coil
Must Operate Voltage
Must Release Voltage
Operate Time
Release Time
Resistance
V
DC
1.25
3.0
2.0
150
4
ms
ms
diode fitted
diode fitted
3.0
2.0
780
Ω
Note. The operate / release voltage and coil resistance will change at a rate of 0.4% per degree C. Values are stated at room temperature (20 degrees C)
Relay
kV
DC or AC peak
17
Isolation Contact/Coil
Insulation Resistance Contact
to all Terminals
10
10
(10
13
)
Ω
min (typical)
Environmental Conditions
ºC
-20 to +70
Operating Temperature Range
*Consult factory for UL ratings
These products have been UL approved for use as per pollution degree 2 classification.
If you require further information as to how this may affect product usage,
please contact c3w_sales@sensata.com
.
Page
2
Copyright © 2022 Sensata Technologies, Inc.
DIMENSIONS
All dimensions are in millimeters.
Page
3
Copyright © 2022 Sensata Technologies, Inc.
Example : DAT71210FU
ORDERING OPTIONS
D
A
T
7
12
10
F
U
Reed Switch Size
Contact Form
A =
n/o
Contact Material
R =
Rhodium
T =
Tungsten
Moulding Ref. No.
Coil Voltage
05 =
5Vdc
12 =
12Vdc
24 =
24Vdc
Isolation Between Contacts
10 =
10kV
75 =
7.5kV
05 =
5kV
Mounting or Connection Style
No suffix indicates PCB mount
F =
PCB mount & coil connection with Flying lead HV connection
P =
Panel mount with wire wrap terminals
S =
PCB mount & coil connection with stud fixing & 1/4” spade HV connection
T =
PCB mount & coil connection with stud fixing & wire wrap HV connection
U
U =
indicates UL approved
Page 4
Made in the UK
Datasheets provided by Sensata Technologies, Inc., its subsidiaries and/or affiliates (“Sensata”) are solely intended to assist third parties
(“Buyers”) who are developing systems that incorporate Sensata products (also referred to herein as “components”). Buyer understands
and agrees that Buyer remains responsible for using its independent analysis, valuation, and judgment in designing Buyer’s systems
and products. Sensata datasheets have been created using standard laboratory conditions and engineering practices. Sensata has not
conducted any testing other than that specifically described in the published documentation for a particular datasheet. Sensata may
make corrections, enhancements, improvements, and other changes to its datasheets or components without notice.
Buyers are authorized to use Sensata datasheets with the Sensata component(s) identified in each particular datasheet. HOWEVER,
NO OTHER LICENSE, EXPRESS OR IMPLIED, BY ESTOPPEL OR OTHERWISE TO ANY OTHER SENSATA INTELLECTUAL PROPERTY
RIGHT, AND NO LICENSE TO ANY THIRD PARTY TECHNOLOGY OR INTELLECTUAL PROPERTY RIGHT, IS GRANTED HEREIN. SENSATA
DATASHEETS ARE PROVIDED “AS IS”. SENSATA MAKES NO WARRANTIES OR REPRESENTATIONS WITH REGARD TO THE DATASHEETS
OR USE OF THE DATASHEETS, EXPRESS, IMPLIED, OR STATUTOR
Y
,
I
NCLUDING ACCURACY OR COMPLETENESS. SENSATA DISCLAIMS
ANY WARRANTY OF TITLE AND ANY IMPLIED WARRANTIES OF MERCHANTABILIT
Y
,
F
ITNESS FOR A PARTICULAR PURPOSE, QUIET
ENJOYMENT, QUIET POSSESSION, AND NON-INFRINGEMENT OF ANY THIRD PARTY INTELLECTUAL PROPERTY RIGHTS WITH REGARD
TO SENSATA DATASHEETS OR USE THEREOF.
CONTACT US
+44 (0)1202 897969
Sensata | Cynergy3
7 Cobham Road,
Ferndown Industrial Estate,
Wimborne, Dorset,
BH21 7PE, United Kingdom
LIABILITY FOR APPLICATIONS ASSISTANCE OR THE DESIGN OF BUYERS’ PRODUCTS. BUYER ACKNOWLEDGES AND AGREES THAT IT
IS SOLELY RESPONSIBLE FOR COMPLIANCE WITH ALL LEGAL, REGULATOR
Y
,
A
ND SAFETY-RELATED REQUIREMENTS CONCERNING
ITS PRODUCTS, AND ANY USE OF SENSATA COMPONENTS IN ITS APPLICATIONS, NOTWITHSTANDING ANY APPLICATIONS-RELATED
INFORMATION OR SUPPORT THAT MAY BE PROVIDED BY SENSATA.
Mailing Address: Sensata Technologies, Inc., 529 Pleasant Street, Attleboro, MA 02703, USA
Copyright © 2022 Sensata Technologies, Inc.
ISO9001CERTIFIED
Rev: 08/22/23